E3000 Critical Point Dryer Pressure Testing
High pressures are generated inside the chamber during the drying process; the
instrument is protected by a safety valve containing a thin diaphragm
guaranteed to rupture at 1850 psi at 20oC.All chambers are pressure
tested to 2500 psi before shipment to the customer. Each chamber bears the
serial number of that test and is supplied with the test certificate.
In some countries local Pressure Vessel Regulations such as the UK Pressure
Systems Safety Regulations 2000 require routine pressure testing and
certification of pressure vessels such as the chamber in the CPD. If this
service is required then it is recommended that the unit is returned to Quorum
Technologies or other competent body for routine pressure testing and
re-certification. Quorum Technologies offer a standard service that includes
seal replacement, pressure testing, re-certification and carriage paid return.
For pricing and further information please
contact us or your local Polaron range
distributor.
ALWAYS REFER TO THE EQUIPMENT OPERATING MANUAL FOR DETAILS OF SPECIFIC SAFETY POINTS FOR THE PRODUCT CONCERNED.