: HOME : ABOUT US : CONTACT US : VISIT US : QUORUM SHOP : VIEW CART : CHECKOUT :

Back to Technical Support    Back to Service Bulletins

E3000 and E3100 Viewing Window Maintenance


Section of the E3000 Series Critical Point Dryer


Viewing Window Maintenance

  1. To remove the viewing window, undo the four hexagon socket screws and remove the clear plastic safety guard.
  2. Unscrew and remove the viewing window slotted retaining ring, being careful not to disturb the end plate
  3. Clean and inspect the retaining ring "O" ring seal. Replace if hardened or damaged.
  4. Remove the viewing window and bonded seal.
  5. Clean and inspect the window and seal. Replace if damaged.
  6. Re-assemble in reverse order, taking care not to damage the window edges.