Back to Technical Support Back to Service Bulletins
E3000 and E3100 Viewing Window Maintenance

Section of the E3000 Series Critical Point Dryer
Viewing Window Maintenance
- To remove the viewing window, undo the four hexagon socket screws and remove the clear plastic safety guard.
- Unscrew and remove the viewing window slotted retaining ring, being careful not to disturb the end plate
- Clean and inspect the retaining ring "O" ring seal. Replace if hardened or damaged.
- Remove the viewing window and bonded seal.
- Clean and inspect the window and seal. Replace if damaged.
- Re-assemble in reverse order, taking care not to damage the window
edges.